← Späť na vyhľadávanie CVE

CVE-2025-15653

Dräger Zeus Infinity Empowered

Popis

Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.

CVSS 6.8EPSS 0.169%Riziko 0.69
Zobraziť zdroj
Zverejnené
2026-06-02 22:16:15
Dotknuté verzie
unknown
Typ
Firmvér
Posledná úprava
2026-07-22 19:10:00
Vektor
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H