CVE-2019-25724

Dräger Infinity M300

Descripción

Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.

CVSS 6.5EPSS 0.17500000000000002%Riesgo 0.66
Ver fuente
Publicación
2026-06-02 20:16:23
Versiones afectadas
cannotmatch
Tipo
Core software
Última modificación
2026-07-22 19:10:00
Vector
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H