CVE-2019-25719

Dräger Infinity Acute Care System

Beschreibung

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

CVSS 8.6EPSS 0.132%Risiko 0.87
Quelle öffnen
Veröffentlicht
2026-06-02 14:16:25
Betroffene Versionen
cannotmatch
Typ
Firmware
Zuletzt geändert
2026-07-22 19:10:00
Vektor
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H