← Zurück zur CVE-Suche

CVE-2019-25718

Dräger Infinity Explorer C700

Beschreibung

Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor.

CVSS 8.4EPSS 0.11800000000000001%Risiko 0.85
Quelle öffnen
Veröffentlicht
2026-06-01 23:16:13
Betroffene Versionen
unknown
Typ
Firmware
Zuletzt geändert
2026-07-22 18:10:00
Vektor
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H